Stanford DS335 Function Generator

3 MHz (Distributed only in Germany.)

Specifications in Brief

Frequency Range
FreqResolution
Sine3.1 MHz1 mHz
Square3.1 MHz1 mHz
Ramp10 kHz1 mHz
Triangle10 kHz1 mHz
Noise3.5 MHz(Gaussian weighting)
Output
Source impedance50 W
GroundingOutput may float up to ± 40 V (AC+DC)
Amplitude
Range50 mVpp to 10 Vpp (50 W), 100 mVpp to 20 Vpp (High Z)
Resolution3 digits (DC offset = 0 V)
Offset± 5 VDC (50 W), ± 10 VDC (High Z)
Offset resolution10 mV (50 W), 20 mV (High Z)
Accuracy0.1 dB (sine output)
Sine Wave
Spurious response<-65 dBc
Harmonic distortion
DC to 20 kHz<-70 dBc
20 kHz to 100 kHz<-60 dBc
100 kHz to 1 MHz<-50 dBc
1 MHz to 3 MHz<-40 dBc
Phase noise<-60 dBc (30 kHz band centered on carrier)
Square Wave
Rise/Fall time (10 to 90 %)15 ± 5 ns
Asymmetry3 ns+1 %
Overshoot<5 %
Ramps and Triangles
Rise/Fall time (10 to 90 %)100 ± 20 ns
Linearity± 0.1 % of full scale
Settling Time200 ns (0.5 % of final value)
FSK Modulation
ModesInternal, External
Max rate50 kHz Internal
External FSKInput sampled every 25 ns
Sweeps
TypeLinear and logarithmic
Type spanLinear (full frequency range), log (6 decades)
Sweep rate0.01 Hz to 1 kHz
Timebase accuracy
Standard± 5 ppm (20 to 30 °C)
OptionalTCXO, 2 ppm stability, 2 ppm aging (20 to 50 °C)
Connectors
Front panel
FUNCFunction output
SYNCTTL level into 50 W
Rear panel
FSK inputTTL
SWP/FSK outTTL
General
InterfacesOptional RS-232 (300 to 9600 Baud, DCE) and GPIB. All instrument functions are controllable over the interfaces.
Non-volatileUp to 9 sets of instrument settings may be stored and recalled
Dimensions8.5 " × 3.5 " × 13 " (W × H × L)
Weight8 lbs
Power22 W, 100/120/220/240 VAC 50/60 Hz